Issue: RT-6100 phoropter is not functioning, or data from pretest devices is not transmitting to MarcoConnect
1. Navigate to the MEM-200 admin console
The black box next to the gray relay box plays an important role in communication between the phoropter unit, the controller, and the PC. It can be rebooted through a web browser. Type this in the address bar in Google Chrome: 192.168.0.120
2. Reboot
The device can be rebooted by pressing the Start button next to the Reboot column at the bottom of the page.
3. Close tab
The page will display "Rebooting", and after ~10 seconds the phoropter will beep. You can close the tab anytime.
4. Resume normal operation
Close and reopen Eyemax/Eyelogic and try controlling the phoropter. Try sending data from the control box.
⭐ Note: If you are still having issues, contact IT Support on Workflow.
If you have any questions you can email ITSupport@for2020now.com
Article Updated on 05/28/2025
Note: This document and any attached materials (or program, etc.) are confidential and proprietary and constitute the sole property of 20/20 Vision Center, LLC and 20/20NOW Eye Exams, PC.. This document and any attached materials (or program, etc.) are intended solely for the recipient and no portion thereof may be reproduced, forwarded, distributed, or otherwise disclosed in any manner to any third party without the express prior written consent of 20/20 Vision Center, LLC and 20/20NOW Eye Exams, PC. © 2017 by 20/20 Vision Center, LLC. All rights reserved.